Modified multiwalled carbon nanotubes paste electrode as a sensor for the electrocatalytic determination of N-acetylcysteine in the presence of high concentrations of folic acid
Authors: Mohsen Keyvanfard, Ali A Ensafi, Hassan Karimi-Maleh, Khadijeh Alizad
Publication date: 2012
Journal name: Analytical Methods
Volume: 4
Pages: 3268-3274
Publisher: Royal Society of Chemistry
A new vinylferrocene modified-multiwalled carbon nanotubes paste electrode was prepared
and used for the determination of N-acetylcysteine (N-AC) and folic acid (FA). Using the
modified electrode, a highly selective square wave voltammetric method for the
simultaneous determination of N-AC and FA was developed. The peak currents of N-AC and
FA increased linearly with their concentration in the ranges of 0.1–662 and 10.0–900 mmol
Lā1, respectively, with corresponding detection limits of 0.09 and 7.0 mmol Lā1. The RSD ...
and used for the determination of N-acetylcysteine (N-AC) and folic acid (FA). Using the
modified electrode, a highly selective square wave voltammetric method for the
simultaneous determination of N-AC and FA was developed. The peak currents of N-AC and
FA increased linearly with their concentration in the ranges of 0.1–662 and 10.0–900 mmol
Lā1, respectively, with corresponding detection limits of 0.09 and 7.0 mmol Lā1. The RSD ...