SPECTROPHOTOMETRIC FLOW INJECTION DETERMINATION OF THIOCYANATE
SPECTROPHOTOMETRIC FLOW INJECTION DETERMINATION OF THIOCYANATE
Ali A Ensafi
Publication date: 1997/4
Journal name: INDIAN JOURNAL OF CHEMISTRY SECTION A
Volume: 36
Pages: 344-346
A spectrophotometric flow injection procedure is described for the determination of
thiocyanate in aqueous medium over the range of 0.005-1.800 mu g/ml. The effect of
reagent concentration, reagent now rates, sample volume, length of the reaction coil and
temperature are reported. The limit of detection is 0.62 ng/ml of thiocyanate. Thiocyanate
can be determined at a rate of 40+/-5 samples/h. The relative standard deviation for ten
replicate analysis of 0.060 mu g/ml of thiocyanate is 1.3%. The system has been applied ...
thiocyanate in aqueous medium over the range of 0.005-1.800 mu g/ml. The effect of
reagent concentration, reagent now rates, sample volume, length of the reaction coil and
temperature are reported. The limit of detection is 0.62 ng/ml of thiocyanate. Thiocyanate
can be determined at a rate of 40+/-5 samples/h. The relative standard deviation for ten
replicate analysis of 0.060 mu g/ml of thiocyanate is 1.3%. The system has been applied ...